10 channels
Up to 10 freely configurable channels, 6 physical and 4 logical, in addition to temperature, which is always present.
A single Centurio PRO measures all the parameters of the plant, up to 10 channels, and controls all the dosing pumps, with dosing proportional to the measurement, residual, volumetric or timed. Changing the probe changes the measured parameter: the same system is used in water mains, in domestic hot water circuits and in wastewater treatment plants.
The probes for the parameters to be controlled are connected to the Centurio PRO. Selecting a probe displays its technical characteristics and the link to the system dedicated to that parameter.
Centurio PRO
Free chlorine0 – 0.5 · 0 – 2 · 0 – 20 mg/l · SCL3N, SCL3S
Total chlorine0 – 2 · 0 – 20 mg/l · SCL8
Chlorine dioxide0 – 0.5 · 0 – 2 · 0 – 20 mg/l · SCL2
Chlorites0 – 2 mg/l · SCT
Bromine0 – 20 mg/l · SBR
Hydrogen peroxide0 – 200 mg/l · SCL9
Peracetic acid0 – 200 · 0 – 2,000 mg/l · SCL11
Ozone0 – 2 · 0 – 20 mg/l · SCL10
Chlorine and chlorine dioxide in hot water0 – 0.5 · 0 – 2 · 0 – 20 mg/l · SCL18, SCL17
pH0 – 14 pH
Redox± 999 mV
Contacting conductivityRange according to the cell constant
Inductive conductivityUp to 0 – 300 mS
TurbidityUp to 0 – 4,000 NTU
Dissolved oxygen0 – 20 mg/l
Corrosion0.001 – 10 MPY
Tracer0 – 300 ppb
Temperature0 – 100 °C · PT100
0-4/20 mA probesRadar level and third-party
Choose a probe: its characteristics and the corresponding system appear below.
0 – 0.5 · 0 – 2 · 0 – 20 mg/l · SCL3N, SCL3S
| SCL3N/0.5 | SCL3N/2 | SCL3N/20 | SCL18/0.5 | SCL18/2 | SCL18/20 | SCL3S/20 | |
|---|---|---|---|---|---|---|---|
| Measuring range | 0 – 0.5 mg/l | 0 – 2 mg/l | 0 – 20 mg/l | 0 – 0.5 mg/l | 0 – 2 mg/l | 0 – 20 mg/l | 0 – 20 mg/l |
| Resolution | 0.001 | 0.001 | 0.01 | 0.001 | 0.001 | 0.01 | 0.01 |
| Free chlorine | Inorganic | Inorganic | Inorganic | Inorganic | Inorganic | Inorganic | Organic and inorganic |
| Water | Fresh | Fresh | Fresh | Fresh | Fresh | Fresh | Fresh or salt |
| Electrodes | 2 | 2 | 2 | 3 | 3 | 3 | 2 |
| Selective membrane | ● | ● | ● | – | – | – | ● |
| Automatic compensation | pH and temperature | pH and temperature | pH and temperature | pH and temperature | pH and temperature | pH and temperature | Temperature |
| Self-cleaning | – | – | – | ● | ● | ● | – |
| Operating pH | 5 – 9.5 | 5 – 9.5 | 5 – 9.5 | 5 – 9 | 5 – 9 | 5 – 9 | 5.5 – 9.5 |
| Operating temperature | 1 – 40 °C | 1 – 40 °C | 1 – 40 °C | 5 – 70 °C | 5 – 70 °C | 5 – 70 °C | 5 – 45 °C |
| Operating pressure | 1 bar | 1 bar | 1 bar | 8 bar | 8 bar | 8 bar | 1 bar |
| Water flow | 30 l/h | 30 l/h | 30 l/h | 40 l/h | 40 l/h | 40 l/h | 30 l/h |
0 – 2 · 0 – 20 mg/l · SCL8
| SCL8/2 | SCL8/20 | SBR/20 | SCL9 | SCL11/200 | SCL11/2000 | SCL10 | SCT | |
|---|---|---|---|---|---|---|---|---|
| Measurement | Total chlorine | Total chlorine | Bromine | Hydrogen peroxide | Peracetic acid | Peracetic acid | Ozone | Chlorites |
| Measuring range | 0 – 2 mg/l | 0 – 20 mg/l | 0 – 20 mg/l | 0 – 200 mg/l | 0 – 200 mg/l | 0 – 2,000 mg/l | 0 – 2 mg/l 0 – 20 mg/l | 0 – 2 mg/l |
| Resolution | 0.001 | 0.01 | 0.01 | 0.1 | 0.1 | 1 | 0.001 0.01 | 0.001 |
| Operating pH | 4 – 12 | 4 – 12 | 6.5 – 9.5 | 1 – 11 | 1 – 9 | 1 – 9 | 2 – 11 | 5.5 – 9.5 |
| Operating pressure | 1 bar | 1 bar | 0.5 bar | 1 bar | 1 bar | 1 bar | 1 bar | 1 bar |
All models: 2 electrodes, selective membrane, automatic temperature compensation, operating temperature 1 – 40 °C, water flow 30 l/h.
0 – 0.5 · 0 – 2 · 0 – 20 mg/l · SCL2
| SCL2/0.5 | SCL2/2 | SCL2/20 | SCL17/0.5 | SCL17/2 | SCL17/20 | |
|---|---|---|---|---|---|---|
| Measuring range | 0 – 0.5 mg/l | 0 – 2 mg/l | 0 – 20 mg/l | 0 – 0.5 mg/l | 0 – 2 mg/l | 0 – 20 mg/l |
| Resolution | 0.001 | 0.001 | 0.01 | 0.001 | 0.001 | 0.01 |
| Electrodes | 2 | 2 | 2 | 3 | 3 | 3 |
| Selective membrane | ● | ● | ● | – | – | – |
| Self-cleaning | – | – | – | ● | ● | ● |
| Operating pH | 4 – 11 | 4 – 11 | 4 – 11 | 5 – 9 | 5 – 9 | 5 – 9 |
| Operating temperature | 1 – 40 °C | 1 – 40 °C | 1 – 40 °C | 5 – 70 °C | 5 – 70 °C | 5 – 70 °C |
| Operating pressure | 1 bar | 1 bar | 1 bar | 8 bar | 8 bar | 8 bar |
| Water flow | 30 l/h | 30 l/h | 30 l/h | 40 l/h | 40 l/h | 40 l/h |
Automatic temperature compensation on all models.
0 – 2 mg/l · SCT
| SCL8/2 | SCL8/20 | SBR/20 | SCL9 | SCL11/200 | SCL11/2000 | SCL10 | SCT | |
|---|---|---|---|---|---|---|---|---|
| Measurement | Total chlorine | Total chlorine | Bromine | Hydrogen peroxide | Peracetic acid | Peracetic acid | Ozone | Chlorites |
| Measuring range | 0 – 2 mg/l | 0 – 20 mg/l | 0 – 20 mg/l | 0 – 200 mg/l | 0 – 200 mg/l | 0 – 2,000 mg/l | 0 – 2 mg/l 0 – 20 mg/l | 0 – 2 mg/l |
| Resolution | 0.001 | 0.01 | 0.01 | 0.1 | 0.1 | 1 | 0.001 0.01 | 0.001 |
| Operating pH | 4 – 12 | 4 – 12 | 6.5 – 9.5 | 1 – 11 | 1 – 9 | 1 – 9 | 2 – 11 | 5.5 – 9.5 |
| Operating pressure | 1 bar | 1 bar | 0.5 bar | 1 bar | 1 bar | 1 bar | 1 bar | 1 bar |
All models: 2 electrodes, selective membrane, automatic temperature compensation, operating temperature 1 – 40 °C, water flow 30 l/h.
0 – 20 mg/l · SBR
| SCL8/2 | SCL8/20 | SBR/20 | SCL9 | SCL11/200 | SCL11/2000 | SCL10 | SCT | |
|---|---|---|---|---|---|---|---|---|
| Measurement | Total chlorine | Total chlorine | Bromine | Hydrogen peroxide | Peracetic acid | Peracetic acid | Ozone | Chlorites |
| Measuring range | 0 – 2 mg/l | 0 – 20 mg/l | 0 – 20 mg/l | 0 – 200 mg/l | 0 – 200 mg/l | 0 – 2,000 mg/l | 0 – 2 mg/l 0 – 20 mg/l | 0 – 2 mg/l |
| Resolution | 0.001 | 0.01 | 0.01 | 0.1 | 0.1 | 1 | 0.001 0.01 | 0.001 |
| Operating pH | 4 – 12 | 4 – 12 | 6.5 – 9.5 | 1 – 11 | 1 – 9 | 1 – 9 | 2 – 11 | 5.5 – 9.5 |
| Operating pressure | 1 bar | 1 bar | 0.5 bar | 1 bar | 1 bar | 1 bar | 1 bar | 1 bar |
All models: 2 electrodes, selective membrane, automatic temperature compensation, operating temperature 1 – 40 °C, water flow 30 l/h.
0 – 200 mg/l · SCL9
| SCL8/2 | SCL8/20 | SBR/20 | SCL9 | SCL11/200 | SCL11/2000 | SCL10 | SCT | |
|---|---|---|---|---|---|---|---|---|
| Measurement | Total chlorine | Total chlorine | Bromine | Hydrogen peroxide | Peracetic acid | Peracetic acid | Ozone | Chlorites |
| Measuring range | 0 – 2 mg/l | 0 – 20 mg/l | 0 – 20 mg/l | 0 – 200 mg/l | 0 – 200 mg/l | 0 – 2,000 mg/l | 0 – 2 mg/l 0 – 20 mg/l | 0 – 2 mg/l |
| Resolution | 0.001 | 0.01 | 0.01 | 0.1 | 0.1 | 1 | 0.001 0.01 | 0.001 |
| Operating pH | 4 – 12 | 4 – 12 | 6.5 – 9.5 | 1 – 11 | 1 – 9 | 1 – 9 | 2 – 11 | 5.5 – 9.5 |
| Operating pressure | 1 bar | 1 bar | 0.5 bar | 1 bar | 1 bar | 1 bar | 1 bar | 1 bar |
All models: 2 electrodes, selective membrane, automatic temperature compensation, operating temperature 1 – 40 °C, water flow 30 l/h.
0 – 200 · 0 – 2,000 mg/l · SCL11
| SCL8/2 | SCL8/20 | SBR/20 | SCL9 | SCL11/200 | SCL11/2000 | SCL10 | SCT | |
|---|---|---|---|---|---|---|---|---|
| Measurement | Total chlorine | Total chlorine | Bromine | Hydrogen peroxide | Peracetic acid | Peracetic acid | Ozone | Chlorites |
| Measuring range | 0 – 2 mg/l | 0 – 20 mg/l | 0 – 20 mg/l | 0 – 200 mg/l | 0 – 200 mg/l | 0 – 2,000 mg/l | 0 – 2 mg/l 0 – 20 mg/l | 0 – 2 mg/l |
| Resolution | 0.001 | 0.01 | 0.01 | 0.1 | 0.1 | 1 | 0.001 0.01 | 0.001 |
| Operating pH | 4 – 12 | 4 – 12 | 6.5 – 9.5 | 1 – 11 | 1 – 9 | 1 – 9 | 2 – 11 | 5.5 – 9.5 |
| Operating pressure | 1 bar | 1 bar | 0.5 bar | 1 bar | 1 bar | 1 bar | 1 bar | 1 bar |
All models: 2 electrodes, selective membrane, automatic temperature compensation, operating temperature 1 – 40 °C, water flow 30 l/h.
0 – 2 · 0 – 20 mg/l · SCL10
| SCL8/2 | SCL8/20 | SBR/20 | SCL9 | SCL11/200 | SCL11/2000 | SCL10 | SCT | |
|---|---|---|---|---|---|---|---|---|
| Measurement | Total chlorine | Total chlorine | Bromine | Hydrogen peroxide | Peracetic acid | Peracetic acid | Ozone | Chlorites |
| Measuring range | 0 – 2 mg/l | 0 – 20 mg/l | 0 – 20 mg/l | 0 – 200 mg/l | 0 – 200 mg/l | 0 – 2,000 mg/l | 0 – 2 mg/l 0 – 20 mg/l | 0 – 2 mg/l |
| Resolution | 0.001 | 0.01 | 0.01 | 0.1 | 0.1 | 1 | 0.001 0.01 | 0.001 |
| Operating pH | 4 – 12 | 4 – 12 | 6.5 – 9.5 | 1 – 11 | 1 – 9 | 1 – 9 | 2 – 11 | 5.5 – 9.5 |
| Operating pressure | 1 bar | 1 bar | 0.5 bar | 1 bar | 1 bar | 1 bar | 1 bar | 1 bar |
All models: 2 electrodes, selective membrane, automatic temperature compensation, operating temperature 1 – 40 °C, water flow 30 l/h.
0 – 0.5 · 0 – 2 · 0 – 20 mg/l · SCL18, SCL17
| SCL3N/0.5 | SCL3N/2 | SCL3N/20 | SCL18/0.5 | SCL18/2 | SCL18/20 | SCL3S/20 | |
|---|---|---|---|---|---|---|---|
| Measuring range | 0 – 0.5 mg/l | 0 – 2 mg/l | 0 – 20 mg/l | 0 – 0.5 mg/l | 0 – 2 mg/l | 0 – 20 mg/l | 0 – 20 mg/l |
| Resolution | 0.001 | 0.001 | 0.01 | 0.001 | 0.001 | 0.01 | 0.01 |
| Free chlorine | Inorganic | Inorganic | Inorganic | Inorganic | Inorganic | Inorganic | Organic and inorganic |
| Water | Fresh | Fresh | Fresh | Fresh | Fresh | Fresh | Fresh or salt |
| Electrodes | 2 | 2 | 2 | 3 | 3 | 3 | 2 |
| Selective membrane | ● | ● | ● | – | – | – | ● |
| Automatic compensation | pH and temperature | pH and temperature | pH and temperature | pH and temperature | pH and temperature | pH and temperature | Temperature |
| Self-cleaning | – | – | – | ● | ● | ● | – |
| Operating pH | 5 – 9.5 | 5 – 9.5 | 5 – 9.5 | 5 – 9 | 5 – 9 | 5 – 9 | 5.5 – 9.5 |
| Operating temperature | 1 – 40 °C | 1 – 40 °C | 1 – 40 °C | 5 – 70 °C | 5 – 70 °C | 5 – 70 °C | 5 – 45 °C |
| Operating pressure | 1 bar | 1 bar | 1 bar | 8 bar | 8 bar | 8 bar | 1 bar |
| Water flow | 30 l/h | 30 l/h | 30 l/h | 40 l/h | 40 l/h | 40 l/h | 30 l/h |
| SCL2/0.5 | SCL2/2 | SCL2/20 | SCL17/0.5 | SCL17/2 | SCL17/20 | |
|---|---|---|---|---|---|---|
| Measuring range | 0 – 0.5 mg/l | 0 – 2 mg/l | 0 – 20 mg/l | 0 – 0.5 mg/l | 0 – 2 mg/l | 0 – 20 mg/l |
| Resolution | 0.001 | 0.001 | 0.01 | 0.001 | 0.001 | 0.01 |
| Electrodes | 2 | 2 | 2 | 3 | 3 | 3 |
| Selective membrane | ● | ● | ● | – | – | – |
| Self-cleaning | – | – | – | ● | ● | ● |
| Operating pH | 4 – 11 | 4 – 11 | 4 – 11 | 5 – 9 | 5 – 9 | 5 – 9 |
| Operating temperature | 1 – 40 °C | 1 – 40 °C | 1 – 40 °C | 5 – 70 °C | 5 – 70 °C | 5 – 70 °C |
| Operating pressure | 1 bar | 1 bar | 1 bar | 8 bar | 8 bar | 8 bar |
| Water flow | 30 l/h | 30 l/h | 30 l/h | 40 l/h | 40 l/h | 40 l/h |
Automatic temperature compensation on all models.
0 – 14 pH
| Electrode | Pressure | Connection | Body | Minimum conductivity | Junction | Temperature | Cable | Diameter | Notes |
|---|---|---|---|---|---|---|---|---|---|
| EpHS | 7 bar (3.5 bar) | BNC | Epoxy | 100 µS | – | 70 °C (80 °C) | 0.8 m | 12 mm | – |
| EpHM | 7 bar (3.5 bar) | BNC | Epoxy | 100 µS | – | 70 °C (80 °C) | 4.5 m | 12 mm | – |
| EpHL | 7 bar (3.5 bar) | BNC | Epoxy | 100 µS | – | 70 °C (80 °C) | 15 m | 12 mm | – |
| EpHSN6 | 7 bar (3.5 bar) | SN6 with PG13.5 thread | Epoxy | 100 µS | – | 70 °C (80 °C) | 1 – 100 m | 12 mm | – |
| EpHMD/100 | 7 bar | BNC | Epoxy | 100 µS | Double | 100 °C | 4.5 m | 12 mm | – |
| EpHM/D | 7 bar (3.5 bar) | BNC | Epoxy | 100 µS | Double | 70 °C (80 °C) | 4.5 m | 12 mm | – |
| EpHM/D/SN6 | 7 bar (3.5 bar) | SN6 with PG13.5 thread | Epoxy | 100 µS | Double | 70 °C (80 °C) | 1 – 100 m | 12 mm | – |
| EpHMD/LI | 7 bar (3.5 bar) | SN6 with PG13.5 thread | Epoxy | 3 µS | Double | 70 °C (80 °C) | 1 – 100 m | 12 mm | For low-conductivity water |
| EpHSC | 7 bar (3.5 bar) | BNC | Epoxy | 100 µS | Double | 70 °C (80 °C) | 4.5 m | 12 mm | Self-cleaning |
| EpHSC/SN6 | 7 bar (3.5 bar) | SN6 with PG13.5 thread | Epoxy | 100 µS | Double | 70 °C (80 °C) | 1 – 100 m | 12 mm | Self-cleaning |
| EpHM/HF | 7 bar (3.5 bar) | BNC | Epoxy | 100 µS | Double | 70 °C (80 °C) | 4.5 m | 12 mm | Resistant to 1% hydrofluoric acid |
| EpHSN6 GK | 10 bar | SN6 | Glass | 150 µS | – | −5 / 135 °C | 1 – 100 m | 12 mm | High temperature |
All models: combination electrode, range 0 – 14 pH, resolution 0.01.
± 999 mV
| Electrode | Range | Pressure | Connection | Body | Minimum conductivity | Junction | Temperature | Cable | Diameter | Notes |
|---|---|---|---|---|---|---|---|---|---|---|
| ERhS | ± 1,000 mV | 7 bar (3.5 bar) | BNC | Epoxy | 100 µS | – | 70 °C (80 °C) | 0.8 m | 12 mm | – |
| ERhM | ± 1,000 mV | 7 bar (3.5 bar) | BNC | Epoxy | 100 µS | – | 70 °C (80 °C) | 4.5 m | 12 mm | – |
| ERhL | ± 1,000 mV | 7 bar (3.5 bar) | BNC | Epoxy | 100 µS | – | 70 °C (80 °C) | 15 m | 12 mm | – |
| ERhHLB SN6 | ± 1,500 mV | 6 bar | SN6 with PG13.5 thread | Glass | 100 µS | – | 80 °C | 10 m | 12 mm | High linearity and sensitivity, for low chlorine concentrations |
| ERhSN6 | ± 1,000 mV | 7 bar | SN6 with PG13.5 thread | Epoxy | 100 µS | – | 100 °C | 1 – 100 m | 12 mm | – |
| ERhMD/100 | ± 1,000 mV | 7 bar | BNC | Epoxy | 100 µS | Double | 100 °C | 4.5 m | 12 mm | – |
| ERhM/D | ± 1,000 mV | 7 bar (3.5 bar) | BNC | Epoxy | 100 µS | Double | 70 °C (80 °C) | 4.5 m | 12 mm | – |
| ERhSC | ± 1,000 mV | 7 bar (3.5 bar) | BNC | Epoxy | 100 µS | Double | 70 °C (80 °C) | 4.5 m | 12 mm | Self-cleaning |
| ERhSC/SN6 | ± 1,000 mV | 7 bar (3.5 bar) | SN6 with PG13.5 thread | Epoxy | 100 µS | Double | 70 °C (80 °C) | 1 – 100 m | 12 mm | Self-cleaning |
All models: combination electrode, resolution 1 mV.
Range according to the cell constant
PVDF body, graphite electrodes.
| Model | Mounting | Connection | Range | K factor | Temperature compensation |
|---|---|---|---|---|---|
| ECDC/1 | In-line / bypass | R3/4" | 0 – 20 mS | 1 | – |
| ECDCC/1 | In-line / bypass | R3/4" | 0 – 20 mS | 1 | NTC 10K |
| ECDCCPT/1 | In-line / bypass | R3/4" | 0 – 20 mS | 1 | PT100 |
| ECDC/10 | In-line / bypass | R3/4" | 0 – 200 mS | 10 | – |
| ECDCC/10 | In-line / bypass | R3/4" | 0 – 200 mS | 10 | NTC 10K |
| ECDCCPT/10 | In-line / bypass | R3/4" | 0 – 200 mS | 10 | PT100 |
| ECDCIM/1 | Immersion | G1/2" | 0 – 20 mS | 1 | – |
| ECDCCIM/1 | Immersion | G1/2" | 0 – 20 mS | 1 | NTC 10K |
| ECDCCIMPT/1 | Immersion | G1/2" | 0 – 20 mS | 1 | PT100 |
| ECDCIM/10 | Immersion | G1/2" | 0 – 200 mS | 10 | – |
| ECDCCIM/10 | Immersion | G1/2" | 0 – 200 mS | 10 | NTC 10K |
| ECDCCIMPT/10 | Immersion | G1/2" | 0 – 200 mS | 10 | PT100 |
Operating temperature 1 – 60 °C (100 °C at 2 bar), operating pressure 7 bar (2 bar at 100 °C).
PVDF body, stainless steel electrodes.
| Model | Mounting | Connection | Range | K factor | Temperature compensation |
|---|---|---|---|---|---|
| ECDI/1 | In-line / bypass | R3/4" | 0 – 5 mS | 1 | – |
| ECDIC/1 | In-line / bypass | R3/4" | 0 – 5 mS | 1 | NTC 10K |
| ECDICPT/1 | In-line / bypass | R3/4" | 0 – 5 mS | 1 | PT100 |
| ECDI/01 | In-line / bypass | R3/4" | 0 – 200 µS | 0.1 | – |
| ECDIC/01 | In-line / bypass | R3/4" | 0 – 200 µS | 0.1 | NTC 10K |
| ECDICPT/01 | In-line / bypass | R3/4" | 0 – 200 µS | 0.1 | PT100 |
| ECDIIM/1 | Immersion | G1/2" | 0 – 5 mS | 1 | – |
| ECDICIM/1 | Immersion | G1/2" | 0 – 5 mS | 1 | NTC 10K |
| ECDICPTIM/1 | Immersion | G1/2" | 0 – 5 mS | 1 | PT100 |
| ECDIIM/01 | Immersion | G1/2" | 0 – 200 µS | 0.1 | – |
| ECDIIM/02 | Immersion | G1/2" | 0 – 500 µS | 0.2 | – |
| ECDICIM/01 | Immersion | G1/2" | 0 – 200 µS | 0.1 | NTC 10K |
| ECDICPTIM/01 | Immersion | G1/2" | 0 – 200 µS | 0.1 | PT100 |
Operating temperature 1 – 60 °C (100 °C at 2 bar), operating pressure 7 bar (2 bar at 100 °C).
Epoxy body, platinum electrodes, diameter 12 mm.
| Model | Mounting | Range | K factor | Temperature compensation |
|---|---|---|---|---|
| ECDHL/1 | In-line / bypass | 0.2 – 20 mS | 1 | – |
| ECDHLC/1 | In-line / bypass | 0.2 – 20 mS | 1 | NTC 10K |
| ECDHLCPT/1 | In-line / bypass | 0.2 – 20 mS | 1 | PT100 |
| ECDHL/01 | In-line / bypass | 0 – 200 µS | 0.1 | – |
| ECDHLC/01 | In-line / bypass | 0 – 200 µS | 0.1 | NTC 10K |
| ECDHLCPT/01 | In-line / bypass | 0 – 200 µS | 0.1 | PT100 |
| ECDHL/10 | In-line / bypass | 20 – 200 mS | 10 | – |
| ECDHLC/10 | In-line / bypass | 20 – 200 mS | 10 | NTC 10K |
| ECDHLCPT/10 | In-line / bypass | 20 – 200 mS | 10 | PT100 |
Operating temperature 1 – 70 °C, operating pressure 7 bar.
Stainless steel body and electrodes, R3/4" connection, in-line or bypass mounting.
| Model | Range | K factor | Temperature compensation | Operating temperature |
|---|---|---|---|---|
| EICDC/1 | 0 – 20 mS | 1 | NTC 10K | 1 – 130 °C |
| EICDCPT/1 | 0 – 20 mS | 1 | PT100 | 1 – 130 °C |
| EICDC/01 | 0 – 200 µS | 0.1 | NTC 10K | 1 – 130 °C |
| EICDCPT/01 | 0 – 200 µS | 0.1 | PT100 | 1 – 130 °C |
| EICDC/001 | 0 – 20 µS | 0.01 | NTC 10K | 1 – 130 °C |
| EICDCPT/001 | 0 – 20 µS | 0.01 | PT100 | 1 – 130 °C |
| EICDCHPT/1 | 0 – 20 mS | 1 | PT100 | 1 – 200 °C |
| EICDCHPT/01 | 0 – 200 µS | 0.1 | PT100 | 1 – 200 °C |
| EICDCHPT/001 | 0 – 20 µS | 0.01 | PT100 | 1 – 200 °C |
Operating pressure 15 bar.
Up to 0 – 300 mS
| ECDINDPT | ECDSINDPT | |
|---|---|---|
| Measuring range | 0.3 – 3 mS 0.3 – 30 mS 0.3 – 300 mS | 0.3 – 10 mS |
| Measuring system | Inductive | Inductive |
| Body | PEEK | PEEK |
| Mounting | In-line / bypass / immersion | In-line / bypass / immersion |
| Temperature compensation | PT100 | PT100 |
| Operating temperature | 1 – 85 °C (100 °C for short periods) | 1 – 85 °C (100 °C for short periods) |
| Operating pressure | 8 bar | 8 bar |
Up to 0 – 4,000 NTU
| ETORB2 | |
|---|---|
| Measuring range | 0/40 NTU – 0/400 NTU – 0/4,000 NTU |
| Measurement error | ± 5% of reading |
| Measuring system | NIR scattered light at 90°, nephelometric principle |
| Body | PVC |
| Optical window | Sapphire glass |
| Automatic compensation | Temperature |
| Self-cleaning | ●(water jet, max 1 bar) |
| Operating temperature | 0 – 40 °C |
| Operating pressure | 2 bar |
0 – 20 mg/l
| EOLUM | |
|---|---|
| Measuring range | 0 – 20 mg/l |
| Resolution | 0.01 |
| Measuring system | Fluorescence |
| Automatic compensation | Temperature and pressure |
| Operating temperature | −5 / +50 °C |
| Operating pressure | 10 bar |
| Body | Stainless steel |
0.001 – 10 MPY
| E-CORR COLD | E-CORR HOT | |
|---|---|---|
| General corrosion | 0.001 – 10 MPY | 0.001 – 995 MPY |
| Localised corrosion | index 0 – 100 | index 0 – 100 |
| Resolution | 0.001 MPY | 0.001 MPY or 1% of reading |
| Conductivity compensation | 10 – 10,000 µS/cm | 10 – 50,000 µS/cm |
| Reading interval | 1, 2, 5 or 20 min | 3 – 1,440 min |
| Alloy factor | 0 – 3 | 0 – 3 |
| Operating temperature | −10 / +50 °C | −10 / +240 °C |
| Operating pressure | 7 bar | 34.5 bar |
| Power supply | 24 VDC, 2 W | 24 VDC, 2 W |
| Outputs | 2 × 0-4/20 mA current loop (general and localised), RS485 | 2 × 0-4/20 mA current loop (general and localised), RS485 Modbus |
| Installation | In-line, NPT threaded connection | In-line, 3/4" NPT or BSP connection |
| Body | AISI 304 | Titanium and PEEK |
| Dimensions | 281.5 mm, Ø 23 / 43 mm | 280 mm, Ø 52 mm |
| Weight | 687 g | 1,150 g |
| Cable | 1.5 m, extensions available | 1.5 m, extensions available |
| Protection | IP65 | IP68 |
0 – 300 ppb
| ETRC2 | |
|---|---|
| Measuring range | 0 – 300 ppb of PTSA |
| Accuracy | ± 1 ppb or 5% of reading |
| Sensitivity | 1 ppb |
| Linearity | 0.99 r² |
| Signal source | LED |
| Detection | Photodiode |
| Response time | 5 seconds |
| Sample pressure | 100 psi (approx. 7 bar) |
| Sample temperature | 4 – 40 °C |
| Ambient temperature | 4 – 49 °C |
| Output | 0-4/20 mA current loop and RS-485 Modbus RTU |
| Power supply | 24 VDC (± 2 V), 65 mA |
| Body | PVC |
0 – 100 °C · PT100
| ETENTC | ETEPT | |
|---|---|---|
| Measuring range | 0 – 100 °C | 0 – 100 °C |
| Resolution | 0.1 | 0.1 |
| Sensor | NTC 10 kΩ | PT100 |
| Pressure | 10 bar | 10 bar |
| Body | PVDF | PVDF |
| Thread | 1/2" | 1/2" |
Radar level and third-party
The probes are installed in flow-through probe holders, in line or by immersion, according to the measuring point of the plant.
Up to 10 freely configurable channels, 6 physical and 4 logical, in addition to temperature, which is always present.
2 ON/OFF set points, 2 proportional set points, one 0-4/20 mA current loop output and one temperature set point.
Thanks to the 0-4/20 mA current loop inputs, third-party measuring probes can be connected, such as pressure transducers, radar level meters or other process instruments; set points and alarm thresholds can be set for each acquired value.
The pump frequency is reduced as the set point is approached. For example, for pH: pump at maximum frequency at 8.00 and stopped at 7.00; for chlorine: maximum frequency at 0.50 mg/l and stop at 1.00 mg/l.
With the amperometric cell on the recirculation, the flow sensor (which stops dosing when there is no circulation) and the temperature probe, it maintains the disinfectant at the set value even when no water is drawn.
With pulse-output water meters, the product concentration, the ppm to be dosed and the volume of each pump injection are set.
Two inputs for pulse-output water meters, for example on inlet and outlet: dosing according to the flow rate and measurement of water volumes.
Water meters →Six timed outputs with weekly or daily timers, for dosing or loads of any kind.
Eight alarm inputs (product levels, valve limit switches…), remote start and stop, input for the flow sensor.
A single Centurio PRO measures free chlorine, turbidity and level and doses chlorine in proportion to the incoming water and to the measured residual.
The system →
The biocide is dosed in proportion to the make-up water and maintained on the recirculation by means of the hot water cells, even when no water is drawn.
The system →
A single Centurio PRO measures dissolved oxygen, pH, Redox, level, flow rate and turbidity and doses flocculant, caustic soda, nutrients, acid, antifoam and chlorine.
The system →The same controller, with software and equipment dedicated to individual applications.
The basic version, for all applications: up to 10 measuring channels, 6 physical and 4 logical, and control of all the dosing pumps of the plant.
Centurio PRO →
For cooling towers: conductivity and bleed-off, inhibitor, two biocides, pH, Redox, tracer and corrosion.
Centurio Tower →
For public swimming pools: free, total and combined chlorine, pH, Redox and temperature; make-up water, drain, dechlorination, flocculant and pool loads.
Centurio POOL →
For domestic hot water: biocide with residual control and proportional to the water meter signal, inhibitor and sampling from the recirculation.
Anti-Legionella →
Inputs, outputs, touch screen, remote control and the probe data sheet are described on the controller page.
Ethernet, Wi-Fi, Modbus RTU, Modbus TCP/IP and current outputs are standard on all Centurio controllers; the 4G modem is available on request. From the web portal it is possible to read the measurements, modify set points, dosing and alarms, receive alarm notifications and download the data; the software is updated online.
Connection to the web portal via network cable: measurements, set points, dosing and alarms accessible from a PC, tablet or smartphone.
Wireless connection to the web portal, with the same functions as the cable connection.
Download of the logged data to a USB memory stick, for archiving and analysis with a spreadsheet.
Web portalPC, tablet and smartphone
Integration into your PLC via serial port.
Integration into your PLC or SCADA via Ethernet network.
Current signal proportional to the measured value, for remote monitoring or for controlling pumps and other instrumentation.
Connection via the 4G network for installations without a local network, with alarm notification by SMS or e-mail.
A single account brings together all registered installations, each with its own devices: the web portal, free of charge and with no subscription, can be accessed from a PC, Mac, tablet or smartphone and allows every parameter to be both viewed and changed, with an encrypted connection and protected access.
On a single screen: probe readings, chemical levels, pump activation thresholds and active alarms, with the status of installations and devices (normal, warning, alarm, standby, not connected).
Installations appear on an interactive map with their status and open with a click; for each one, the local weather conditions and temperature are also shown.
Archive of measurements and alarms, available for consultation at any time with charts and numerical reports exportable in Excel format.
Alarm thresholds for readings, dosing, faulty probe and out-of-range readings, with notification by e-mail or SMS depending on the communication system; a warning is also issued when an installation remains disconnected beyond the set time.
Remote modification of the set points, of the dosing pump activation thresholds and of the corresponding hourly flow rate.
Status of relays, pulse outputs, digital inputs and current outputs, with the name assigned to each connection.
Tasks are scheduled by installation or by device and assigned to technicians; a reminder is sent by e-mail two days before the due date. Each task is shown as scheduled, due, overdue or completed.
The main account has full access and creates the technicians’ accounts, each enabled only for the assigned installations and devices; the last login and activity of each technician are visible.
Reports on devices, tasks and technicians, downloadable in PDF or CSV with the customer’s logo; the log keeps the history of operations and disconnections.
Logo, main colour and light or dark theme; the tiles on the main screen can be rearranged by dragging them. Two installations can be displayed side by side for comparison.
Each device is added through a guided procedure: the model is recognised automatically and associated with the installation, together with the technician to be contacted.
As an alternative to the portal, the Modbus RTU or TCP/IP port allows integration into the plant supervision system.
Tell us the parameters to be controlled and the characteristics of the plant: we will propose the most suitable probes, pumps and accessories.
Call or write to us: a technician will reply with a tailored proposal of controllers, pumps, probes and accessories.