Colour display
Multicolour backlit display that indicates the different operating states by changing colour. Programming via a single knob.
The LDsCL controller for ozone measures the ozone dissolved in the water by means of the SCL10 closed amperometric cell, with selective membrane and automatic temperature compensation. Since ozone is generated on site and decomposes rapidly, the controller starts and stops the generator until the set point is reached.
It acquires the measurement from the cell and controls the dosing pump in proportion to the measured value.
Multicolour backlit display that indicates the different operating states by changing colour. Programming via a single knob.
Proportional, ON/OFF, PWM and PID dosing, with delayed dosing start.
Permanent data storage and a system log viewable on the display; the internal log can be exported to a USB memory stick for processing with a spreadsheet.
Chemical level, no flow, maximum dosing time exceeded, probe failure, out-of-range reading; audible or visual signalling via the alarm output.
By changing colour, the controller’s backlit display indicates the different operating states, which can be recognised even from a distance.
Measurement and dosing according to the set parameters.
Controller in stand-by.
Warning signal, for example the chemical level.
Alarm signal, for example from the probe.
The instrument is available in two versions, which differ in the number of inputs and outputs: the PLUS version adds the meter input, PID control, a second level input, a second dosing pump output, the 0-4/20 mA current loop outputs and the outputs for probe self-cleaning and for the circulation pump.
| BASIC | PLUS | |
|---|---|---|
| Measuring channels | 1 + temperature | 1 + temperature |
| Level inputs | 1 | 2 |
| Flow input | 1 | 1 |
| Stand-by input | 1 | 1 |
| Meter input | – | 1 |
| Outputs for pumps and loads (relays) | 1 (ON/OFF, PWM) | 2 (ON/OFF, PWM, PID) |
| Proportional outputs for dosing pumps | 1 | 2 |
| Alarm output | 1 | 1 |
| 0-4/20 mA current loop outputs | 2 on request | 3 |
| Probe self-cleaning output | – | 1 |
| Circulation pump output | – | 1 |
| Modbus RTU | ● | Optional |
| Ethernet, Wi-Fi, 4G, USB | Optional | Optional |
Inputs, outputs, control modes and alarms of the controller.
For measuring ozone in water supply networks, drinking water treatment, the food industry and swimming pools. Ozone is generated on site and rapidly decomposes into oxygen: the controller starts and stops the ozone generator until the set point is reached.
SCL closed or ECL open amperometric cells are connected to the LDsCL controller, resulting in the different versions.
Learn more: Redox, open amperometric cell and closed amperometric cell →
The electrodes are protected by a selective membrane that allows only the oxidant of interest to be measured, without interference from other substances. Automatic temperature compensation (and pH compensation for free chlorine), resolution down to 0.001 mg/l.
SCL cells →
Membrane-free, with 3 electrodes (one measuring and two reference), with ball self-cleaning: suitable up to 70 °C and less sensitive to fouling. For free chlorine (SCL18) and chlorine dioxide (SCL17).
SCL17 and SCL18 →
Direct measurement of free chlorine in mg/l in swimming pool water, including salt water.
ECL cells →The cell consists of an electronic section and a galvanic section.
Learn more: Redox, open amperometric cell and closed amperometric cell →The measuring electrodes are protected from direct contact with the water by a semi-permeable membrane: only the specific oxidant is measured, without interference from other substances that could alter the measurement.
The electrodes of the galvanic section detect the electric current generated by the dissociation of the oxidant.
The cell electronics compensate the reading for temperature (and, where applicable, pH) and send the signal to the controller, which shows the concentration in mg/l on the display.
Measuring range, resolution and characteristics of the cell for measuring ozone. The controller is common to all parameters: inputs, outputs and connections are described in the controller sections.
| Parameter | Range | Resolution | Cell | Compensation |
|---|---|---|---|---|
| Ozone | 0 – 2 mg/l 0 – 20 mg/l | 0.001 mg/l 0.01 mg/l | SCL10 | Temperature |
| Temperature | 0 – 100 °C | 0.1 °C | – | – |
| Characteristic | Value |
|---|---|
| Average power consumption | 25 W |
| Relay outputs | 5 A at 230 VAC, fuse-protected |
| Delayed dosing start | Up to 60 minutes |
| Networking | Up to 31 controllers |
| Ambient temperature | −10 – 50 °C |
| Relative humidity | 0 – 95 %, non-condensing |
| Protection rating | IP65 |
| Housing | ABS |
| Weight | 1.45 kg |
| Mounting | Wall-mounted, with four fixing points |
The controller is the core of the measuring and dosing system: for each application, the system page gives the composition, the installation diagram and the installation.
Control of residual ozone with control of the generator.
The system →
Prevention guide: biofilm, regulatory framework, comparison of disinfectants and systems for domestic hot water.
The sector →For wastewater and for plants with special requirements the system is defined on the characteristics of the plant: contact us.
For measuring ozone in water supply networks, drinking water treatment, the food industry and swimming pools. Ozone is generated on site and rapidly decomposes into oxygen: the controller starts and stops the ozone generator until the set point is reached.
Closed amperometric cell with selective membrane.
Cell details →| SCL10 | |
|---|---|
| Measuring range | 0 – 2 mg/l 0 – 20 mg/l |
| Resolution | 0.001 0.01 |
| Operating pH | 2 – 11 |
| Operating pressure | 1 bar |
| Electrodes | 2 |
| Selective membrane | ● |
| Automatic compensation | Temperature |
| Operating temperature | 1 – 40 °C |
| Water flow | 30 l/h |
| Response time (T90) | approx. 50 s |
| Start-up time | approx. 1 h |
| Replacement electrolyte | ELESCL10 |
| Replacement membrane | MESCL10 |
| Materials | PVC-U and 1.4571 stainless steel |
| Dimensions | Ø 25 × 241 mm |
| Maintenance | Electrolyte replacement every 3 – 6 months |
The controllers support two-way remote access (read and write) through the web portal from a PC, tablet or smartphone: measurements can be viewed, set points, dosing and alarms changed, alarms received and logs downloaded. Alternatively, they can be integrated into your PLC via Modbus.
Connection to the web portal via network cable: measurements, set points, dosing and alarms accessible from a PC, tablet or smartphone.
Wireless connection to the web portal, with the same functions as the cable connection.
Connection via the 4G network for installations without a local network, with alarm notification by SMS or e-mail.
Web portalPC, tablet and smartphone
Integration into your PLC via serial port.
Current signal proportional to the measured value, for remote monitoring or for controlling pumps and other instrumentation.
Download of the logged data to a USB memory stick, for archiving and analysis with a spreadsheet.
One communication type at a time: Ethernet, Wi-Fi, 4G or Modbus RTU (built into the BASIC versions, optional on the PLUS versions). 0-4/20 mA current loop outputs standard on the PLUS versions, optional on the BASIC versions; meter input on the PLUS versions only.
A single account brings together all registered installations, each with its own devices: the web portal, free of charge and with no subscription, can be accessed from a PC, Mac, tablet or smartphone and allows every parameter to be both viewed and changed, with an encrypted connection and protected access.
On a single screen: probe readings, chemical levels, pump activation thresholds and active alarms, with the status of installations and devices (normal, warning, alarm, standby, not connected).
Installations appear on an interactive map with their status and open with a click; for each one, the local weather conditions and temperature are also shown.
Archive of measurements and alarms, available for consultation at any time with charts and numerical reports exportable in Excel format.
Alarm thresholds for readings, dosing, faulty probe and out-of-range readings, with notification by e-mail or SMS depending on the communication system; a warning is also issued when an installation remains disconnected beyond the set time.
Remote modification of the set points, of the dosing pump activation thresholds and of the corresponding hourly flow rate.
Status of relays, pulse outputs, digital inputs and current outputs, with the name assigned to each connection.
Tasks are scheduled by installation or by device and assigned to technicians; a reminder is sent by e-mail two days before the due date. Each task is shown as scheduled, due, overdue or completed.
The main account has full access and creates the technicians’ accounts, each enabled only for the assigned installations and devices; the last login and activity of each technician are visible.
Reports on devices, tasks and technicians, downloadable in PDF or CSV with the customer’s logo; the log keeps the history of operations and disconnections.
Logo, main colour and light or dark theme; the tiles on the main screen can be rearranged by dragging them. Two installations can be displayed side by side for comparison.
Each device is added through a guided procedure: the model is recognised automatically and associated with the installation, together with the technician to be contacted.
As an alternative to the portal, the Modbus RTU port allows integration into the plant supervision system.
Tell us the product to be dosed, the volumes of water treated and the characteristics of the plant: we will propose the most suitable cell, pump and accessories.
Call or write to us: a technician will reply with a tailored proposal of controllers, pumps, probes and accessories.